Replacing fuses – PHS/Multi PHS
- If the PHS or Sandblaster does not turn on check the fuses. You will find all the fuses within the Service access plate.
- Turn off the PHS and check the main fuse:
- – PHS 110 V – unscrew the 16A fuse. Check if the fuse is blown. If so, replace it with a new one.
- – PHS 230 V – Check if the residual-current device (RCD) is switched off. If so, switch it on.
- Turn on the PHS and check if it works. If not, turn off the device and unscrew the Service access to the fuse panel. Use the 5 mm Allen key.
- Check if the low voltage fuse (6.3 A) is not blown. If so, replace it with a new one.
- Turn on the device and check if it works.
- If the PHS is working and the Sandblaster is not, turn off the PHS.
- Check the Sandblaster fuses. If the Sandblaster fuses (1 A) are blown, replace them.
- Turn on the PHS and check if it works.
- If there is still a problem, contact our After-Sales team at [email protected].
- Dispose of the used fuse according to your local waste management policy.